Measurement of Low Concentration of Nanosized Objects Suspended in a Liquid Medium

Thumbnail
View/Open
Measurement-of-Low-Concentration-of-Nanosized-Objects.pdf (616.6Kb)
Author
Merkulovs, Dmitrijs
Vilitis, Oskars
Kozlovs, V.
Date
2018
Metadata
Show full item record
Abstract
The new optical scheme of refractometer with temperature stabilisation 10-2 °C is developed, which allows measuring a refractive index of the sample with accuracy not worse than 10-5; dependence of refraction index on concentration of SiO2 nanoparticles in liquid suspension is obtained within the framework of the research.
URI
https://dspace.lu.lv/dspace/handle/7/52449
DOI
10.2478/lpts-2018-0015
Collections
  • Zinātniskie raksti (CFI) / Scientific articles [604]

University of Latvia
Contact Us | Send Feedback
Theme by 
@mire NV
 

 

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

Login

Statistics

View Usage Statistics

University of Latvia
Contact Us | Send Feedback
Theme by 
@mire NV